LuciPac Pen    Prodotto

PREPARATION FOR SWAB TEST

  • Remove the cotton swab stick (orange holder) from the main tube. Then, moisten the tip of the cotton swab with tap water.
  • Swab the object.
  • Put the swab stick back into the main tube. And then push the swab stick downwards to break the capsule in the tube.
  • Shake the whole LuciPac Pen tube repeatedly so that all remaining liquid in the capsule falls into the reaction tube and dissolve the reagent in the tube.
  • lnsert the whole LuciPac Pen tube into the measurement chamber of the Lumitester PD-20. And close the measurement chamber cover.
  • Press ENTER key on the operation panel to start the measurement. Measurement result is displayed after countdown.