LuciPac LS3.2-400 is a kit for testing cleanliness levels
mainly for canalicular objects using bioluminescence techniques
using firefly luciferase developed with Kikkoman's unique
biotechnology.
This kit contains specialized long stem swabs suitable for
swabbing inside of the channels of medical endoscopes and
of filling nozzles in beverage industry, which have not been
accessible by LuciPac Pen and/or other ATP swab test kits.
Size of the long stem swab: Cotton bud 3.2mm Ø and stem length
400 mm
Firefly luciferase emits light in the presence of ATP and
luciferin. The AMP produced from this reaction is converted back
into ATP using PPDK to enable a high but stable amount of
luminescence to be obtained.
ATP is a source of energy necessary for various forms of life
that is present in organic residues, such as microorganisms, food
residue, and biological substances that originate from other living
organisms. This method allows you to measure and detect organic
This kit can be used to test cleanliness levels using a swab test.
Do not use this device for any purpose other than what it was
designed for as a tester of cleanliness levels. Please be aware that
this kit cannot be used to test or measure for counts of viable
bacteria or more specific types of pathogenic bacteria.
This kit is used to indicate cleanliness levels in food production
centers and medical treatment facilities. This kit shall not
provide any guarantee that a given test environment is free of
bacterial contamination. This kit cannot be used to test food or
food products directly. This kit is not a medical device.


How to use

Remove the LuciPac Pen-AQUA from refrigerator, and wait
until they have reached room temperature (see Table 2).Use them
as soon as possible once they have returned to room temperature.
Wear lab globes to prevent contamination and take out the
long stem swab carefully so as to not to touch the parts as this
may affect the cleanliness level from the end of the stem of the
individual package.
If running the measurement test on a dry surface, first moisten
the cotton bud or the test subject with tap water. Please
avoid using saline solution instead of water.
Handle the long stem swab with care so as to not to touch the
parts as this may affect the cleanliness level.Insert the long
stem swab in the canalicular object to be tested, and wipe it off
as long as the length of the long stem swab reaches, and pull
the swab up slowly.
Hold the long stem swab at 12 cm or 13 cm distant point from
the cotton bud, and remove the sampling stick from the main
body (casing) of the LuciPac Pen-AQUA by your free
fingers.Handle the sampling stick with care so as to not to
touch the parts as this may affect the cleanliness level.
Insert the long stem swab in the main body (casing) of the
LuciPac Pen-AQUA to push and break the center of the upper
seal of the container for releasing reagent. Be careful not to
break bottom seal of the container for releasing reagent. Wash
the cotton bud in the releasing reagent. Remove the long stem
swab after thoroughly washing the cotton bud in the releasing
reagent.
Return the sampling stick to the main body (casing) and push
it all the way into the main body (casing).
Hold firmly onto the LuciPac Pen-AQUA casing and shake it.
Allow leftover luminescent reagent to throughly dissolve.
Insert the LuciPac Pen-AQUA into the Lumitester and close
the Lumitester measurement chamber cover.