LuciPac W for Lumitester PD 10 N    Product

A contamination is measured as follows:

1. Wipe the to-be-tested surface with the sterile LuciPac W cotton swab.
2. Put the swab back in the tube and push the swab stick until it has opened the reagent capsule. To do so, place the tube unit on the table or hold it by hand.
3. Shake the LuciPac W several times so that all the liquid flows into the reagent capsule.
4. Insert the LuciPac W into the measuring chamber of the Lumitester PD-10 N and close the lid. Press -Enter -. After 10 seconds, the result will be displayed.